CALL US

02 9543 7377

Interface 1000

The Interface 1000 is a perfect potentiostat/galvanostat/ZRA for use in general electrochemistry applications

Product Category: Potentiostats

Industry: Electrochemistry

Supplier: Gamry

Description

Overview

Flexible
With 9 current ranges and 3 gain stages, you have 11 decades of current measurement at your command. From corrosion to batteries, sensors to supercaps, the Interface 1000 has got you covered.

Affordable
By combining state of the art electronics with expert design, the Interface 1000 is quite simply the best value in potentiostats.

Powerful
Cycle your batteries. Polarize those large electrodes. The Interface 1000 delivers a full 1 Amp of current up to the compliance limit of ±20 V.

Floating
Like all Gamry Potentiostats, the Interface 1000 was designed from the ground up for true floating capability. Whether you need to measure earth grounded working electrodes, or multiple electrodes in a shared electrolyte, the Interface 1000 is up to the task.

Low Noise
The Interface 1000 continues the Gamry tradition with an exceptional < 20 μV noise specification. Low noise = better measurements.

Impedance on Board
No external modules to add - simply purchase a software license and accurately acquire impedance measurements up to 1 MHz.

Configurable Multichannel
Why sacrifice convenience when you need more than one potentiostat? The Interface 1000 lets you control where each channel is located. Put channel 1 in your hood, channel 2 in your glovebox, and dock channels 3-8 on the bench. You wanted flexibility, we delivered!

Accuracy Contour Plot

The Accuracy Contour Plot provides a detailed look at the performance you can expect from your instrument in real-world situations. The results include the cell cable.

Interface 1000 Accuracy Contour Plot

Specifications

Interface 1000
Potentiostat YES
Galvanostat YES
Zero Resistance Ammeter YES
Cell Connections 2, 3, or 4
Floating (Isolated from Earth) YES
WEIGHT 2.4 kg
DIMENSIONS 24 (W) x 6 (H) x 27 (D) cm
SYSTEM
Max Current ±1 A
Current Ranges 9 (10 nA - 1 A)
Current Ranges (w/Internal

Gain applied)

11 (100 pA - 1 A)
Min Voltage Resolution 1 μV
Min Current Resolution 3.3 fA
Max Applied Potential ±12 V
Rise Time TBD
Noise and Ripple <20 μVrms
Min Time Base 10 μs
Max Time Base 715 s
Min Potential Step 12.5 μV
Analog/Digital Converter 16 bit
EIS MEASUREMENT
Frequency Range 10 μHz - 1 MHz
EIS Accuracy See Accuracy

Contour Plot

Max AC Amplitude 2.33 Vrms
Min AC Amplitude 17.8 μVrms
CONTROL AMP
Compliance Voltage ±20 V
Output Current >±1 A
Speed Settings 5
Unity Gain Bandwidth 980, 260, 40, 4, 0.4 kHz
ELECTROMETER
Input Impedance >1012 Ω
Input Current <20 pA
Bandwidth (-3dB) (typical) >10 MHz
Common Mode Rejection Ratio >80 dB (10 kHz)

>60 dB (1 MHz)

APPLIED POTENTIAL
Accuracy ±1 mV ±0.2% of setting
Resolution 12.5 μV, 50 μV, 200 μV/bit
Drift <20 μV/°C
Potential Scan Range ±0.4 V, ±1.6 V, ±6.4 V
MEASURED POTENTIAL
Accuracy ±1 mV ±0.3% of reading
Full-Scale Ranges ±12 V, ±3 V, ±300 mV, ±30 mV
Resolution 400 μV, 100 μV, 10 μV, 1 μV/bit
Offset Range ±12 V, ±3 V
APPLIED CURRENT
Accuracy ±5 pA ±0.3% of range
Resolution 0.0033% full-scale/bit
MEASURED CURRENT
Accuracy ±5 pA ±0.3% of range
Resolution 0.0033% full-scale/bit
Bandwidth (-3 dB)

NOTE: Bandwidth is current

range dependent.

> 10 MHz (1A - 100 μA)

> 1.5 MHz (10 μA)

> 0.15 MHz (1 μA)

Stability Settings 3
Post Offset Gain 1X, 10X, 100X
Offset Range ±1X full-scale
iR COMPENSATION
Mode Current Interrupt
Min Interrupt Time 33 μs
Max Interrupt Time 715 s

 

Scientific Solutions