Adsorbate Such as Carbon monoxide, ammonia and substances that are liquid at room temperature. (Except corrosive/ hyper-toxic/hazardous gas) Vapor sorption temperature 0 - 50 ℃ Pore size distribution range 0.35 - 500 nm

Product Category: Vapour Sorption

Industry: Life Sciences, Membrane Manufacture and Testing, Particle and surface science, Pharmaceutical, Personal Care and Natural Products

Supplier: JWGB

Description

Preparation of adsorbate by evaporator.
Use stainless steel and thermostatic control module as manifold. Constant temperature range: room temperature to 50 oC.
Effective cold trap system prevents vapor from polluting vacuum pump and environment.
FFKM (Perfluoroelastomer) O rings for banner sealing performance.
Applications
Measurements of BET surface area and pore size distribution.
Measurements of various gas and steam adsorption isotherms.
Testing of isosteric heat of adsorption.
Determination of selective adsorption coefficient.

Specifications

Parts Vapor 200C Vapor 100C Vapor 100B
Testing Principle The static volumetric method
Data Features BET surface area: ≥ 0.01 m2/g
Mesopore: 2 - 500 nm (BJH)
Micropore: 0.35 - 2 nm (t-plot, DR, MP,DA, HK ,SF)
Physical Adsorbate N2, Ar, Kr, CO2
Vapor Adsorbate Carbon monoxide, ammonia, and a substance that is liquid at
room temperature.(Except corrosive, hyper-toxic, hazardous gas) Same as Vapor 200C and Vapor 100C,
but p0 is greater than 1KPa
Physical Adsorption
Temperature 77K, 87K
Vapor Adsorption Temperature 0 - 50 oC
Steam Recovery Cold trap in analysis station Double cold trap in analysis and pretreatment station
Pretreatment Vacuum in-situ pretreatment Separate preposition and vacuum in-situ pretreatment
Essential Parameters
Evaporator Phase of adsobate transforming from liquid to gas by evaporator in room temperature, implements test requirements
Adsorption Pressure P/P0 is less than 1 (P0 is saturated vapor pressure of adsobate in adsorption temperature)
Vacuum Pump Special corrosion – resistant two stage rotary vane vacuum pump, molecular pump
Pressure Sensor 1000 Torr, 10 Torr, 1Torr
Sorption Temperature -196 oC is provided by liquid nitrogen in low temperature Dewar flask; 0 - 50 oC is provided by thermostatic water bath.
Number of Gas Inlet Providing separate gas intakes according to customer requirements. (Number of intakes is multiples of 4).
Gas Recovery Effective cold trap system prevents from polluting vacuum pump and environment by vapor.
Pretreatment Vacuum in –situ pretreatment, temperature ≤400 oC ±1 oC.

Model comparison

Parts                                         Vapor 200C                     Vapor 100C                                              Vapor 100B
Evaporators                                 1                                                1                                                                1
Vapor Sorption Station              1 (1/0.1 Torr)                  1 (1/0.1 Torr)                                     1 (10 Torr)
Micropore Analysis Station          1 (1/0.1 Torr)                   NO                                                       NO
Pressure Sensors                              7                                        5                                                           4
Cold Trap                                          1                                       2                                                           2
Preparation Bit                   NO( Can be prepared in situ)    1                                                             1
Thermostatic Water Bath           1 (0 - 50 oC)                   1 (0 - 50 oC)                               1 (0 - 50 oC)
Dewar flask (1L)                                  2                                   NO                                                  NO
Dewar flask (3L)                                  1                                       2                                                      2
Mechanical Pump One corrosion-resisting pump One corrosion-resisting pump,and One mechanical pump One corrosion-resisting pump(or One mechanical pump)
Molecular Pump                             1                                          1                                                            NO