Quick-200 multi-stations surface area analyzer
Product Category: Surface Area
Industry: Particle and surface science
Supplier: AMI
Description
Hardware and Operation
Basic performance parameters
Specific surface area: >0.0001m2/g
Pore size: 2nm -500nm
Aperture resolution: 0.02nm
Power: 110~220V, 50/60Hz

TECHNIQUES
- Isotherm curve
- BET (Brunauer, Emmett, and Teller) surface area
- Langmuir surface area
- t-Plot
- Alpha-s method
- The BJH (Barrett, Joyner, and Halenda) method
- Dollimore-heal adsorption and desorption curve
- Horvath-Kawazoe (H-K)
- The MP method
- DFT
- NLDFT
- Dubinin-Astakhov (D-A)
- Dubinin-Radushkevich (D-R)
- Summary reports
- Custom reports

CAPABILITIES
- Physisorption
- Gas adsorption
- Gas sorption
- Nitrogen sorption
- Surface characterization
- Pore analysis
- Pore size
- Pore size distribution
- BET surface area

Specifications
|
|
QUICK200 |
||
|
Analysis port(s) |
4 |
8 |
|
|
p0 Port(s) |
1 |
2 |
|
|
Number of micropore analysis stations |
NONE |
NONE |
|
|
Adsorbates |
N2 , Ar, Kr, H2 , O2 , CO2 , CO, NH3 , CH4 , etc. |
||
|
Surface Area Range |
0.0001 m2/g to unknown upper limit; Measurement accuracy (standard sample) ≤ ± 1.0 % |
||
|
Pore Size Range |
2nm – 500 nm(Options)
Repeatability: ≤ 0.2 nm in mesopore range and ≤ 0.02 nm in micropore range |
||
|
Minimum Pore Volume |
0.0001 cm3/g |
||
|
Pressure Sensors (Individual sensor systems for each analysis port can be selected) |
Four high precision 1000torr sensors for analysis ports. extra 1000 Torr for p0 determination |
||
|
Pressure Sensor Accuracy |
± 0.15 % (Full Scale) |
||
|
Degassing Stations |
4 insitu |
Standard independent degassing station |
|
|
Degassing Temperature |
Room temperature to 400 °C (optional 500 °C), accuracy 1 °C |
||
|
Vacuum System |
Standard mechanical oil pump (ultimate vacuum 103 Pa) |
||
|
Independent vacuum system for each analysis port |
√ (Options) |
NONE |
|
|
Dimensions |
|
|
|
|
Temperature Requirements of Environment |
15 – 40 °C |
||
|
Humidity Requirements of Environment |
10 % – 90 % |
||
|
Power Requirements |
AC 220 V ± 20 V, 50/60 Hz, maximum power 300 W, current 5 A |
||
|
Steam absorption |
× |
× |
|
|
Independent Intake Gas System for Each Analysis port |
√ (Options) |
NONE |
|