EIS Box
Make impedance measurements on up to eight batteries
The EIS Box™ is a multiplexed eight channel instrument designed for impedance measurements on batteries (or other devices such as supercapacitors or fuel cells) and is typically used in conjunction with a battery cycler. It is isolated from ground and is capable of EIS from 100 kHz down to 10 µHz and comes in a 3U 19" rack-mountable chassis.- EIS on up to eight cells
- 10 µHz to 100 kHz*
- Includes 3 m cables
- Ethernet based
- Up to 6V per channel
Product Category: Potentiostats
Industry: Electrochemistry
Supplier: Gamry
Description
The EIS Box comes with eight 3 m cell cables and is designed to be used in conjunction with a battery tester/cycler to provide impedance testing. Cycle testing of the battery would be provided by the battery tester/cycler. It can be installed in a 19-inch rack (similar to the ECM8 multiplexer). Connection to the EIS Box is made through ethernet or USB. As with a USB connection to our potentiostats, the EIS Box has to be in constant contact with the host computer for experiments to be performed successfully. Connection via ethernet is done via a direct connection to the computer or over a local network only; connection over the internet is not possible.
Specifications
SPECIFICATIONS |
|
Cell Connections | Four |
Number of Cells | Up to Eight |
Cables | Set of Current and Sense Cables – 3 m standard length |
DC Power Input Voltage | 100-240, 50-60 Hz, 2.1A |
Connection to computer | Ethernet or USB |
EIS Measurement | 10 uHz to 100 kHz* (*specifications subject to change) |
Max AC Amplitude | 3 V or 5 A |
Accuracy | See Accuracy Contour Plot |
Weight | Approximately 25 lb (10 kg) |
Dimensions | 5.25”x18.875”x18.875” (13.3 cm H x 41.5 cm W x 45.7cm D) |
Power | Max 120W |
Mounting | Fits 19” (48.3 cm) standard rack mount 3 U height |
Fuses | IEC 127 For 115 VAC, use 6.3 A 250 V For 230 VAC, use 3.153 A 250 V |
CONTROL AMPLIFIER |
|
Compliance Voltage | Min. ±6.5 V |
Output Current | Min. ±5 A |
Unity Gain Bandwidth | 1050, 250, 43, 4.4, 0.5 kHz |
Slew Rate | 10, 4.4, 0.7, 0.07, 0.007 V/µs |
DIFFERENTIAL ELECTROMETER |
|
Max Input Voltage | Max ±6.0 V |
Input Current | Max 25 pA |
Input Resistance
Differential (between inputs) |
10 TΩ |
Input Capacitance
Differential (between inputs) |
8 pF |
Bandwidth (–3 dB) | Min 12 MHz |
CMRR
DC to 10 kHz |
98 dB |
VOLTAGE MEASUREMENT |
|
A/D Full Scale Ranges | ±6.5536 V |
Resolution (before gains) | 200 µV/bit |
Zero Offset Error | 0.5 mV |
Gain Error | 0.2% |
Offset Range | ±6.5536 V |
Post-offset Gain | 1×, 10×, 100× |
CURRENT TO VOLTAGE CONVERTER |
|
Maximum Full-scale Range | ±5 A |
Minimum Full-scale Range | ±50 µA ±500 (after 100× gain) nA |
Voltage across Rm | ±100 mV at full scale |
Input Offset Current | Max 25 pA |
Range Zero Offset | Max 0.05% of range |
Gain Tolerance, 5 A to 50 µA ranges | Max 0.2% of range |
Bandwidth (-3 dB) | 5 MHz |
CURRENT MEASUREMENT |
|
Resolution | 0.00333…% FS/bit |
Offset Range | ±100% of range |
Post-offset Gain | 1×, 10×, 100× |
Accuracy | Dominated by current-to-voltage error (see above) |
ENVIRONMENTAL |
|
Operating Temperature Range | 0 to +45°C |
Relative Humidity | Max 90% (non-condensing) |
Storage and Shipping Temperature | –25 to +75°C |
Maximum Shipping Acceleration | 30g |