Saccharimeter Autopol 880 & 880T

Recommended for the sugar industry, the Autopol 880 saccharimeter is a dual wavelength microprocessor based digital saccharimeter capable of measuring at both 589 nm and 880 nm. The 880 nm ICUMSA Approved near infrared wavelength allows measurement of dark sugar factory juices, while the AUTOPOL Automatic Saccharimeter Purity Mode allows simultaneous measurement and display of PURITY, BRIX and POL.

Product Category: Saccharimeters

Industry: Food, Dairy and Wine

Supplier: Rudolph Research Analytical

Description

AUTOPOL 880 Series – Digital Saccharimeter

Recommended for the sugar industry, the Autopol 880 saccharimeter is a dual wavelength microprocessor based digital saccharimeter capable of measuring at both 589 nm and 880 nm. The 880 nm ICUMSA Approved near infrared wavelength allows measurement of dark sugar factory juices, while the AUTOPOL Automatic Saccharimeter Purity Mode allows simultaneous measurement and display of PURITY, BRIX and POL.

Dual Wavelengths

The most important feature of the AUTOPOL® saccharimeter models IIZ, 880 and 880T are their ability to measure at two wavelengths: 589 nm and 880 nm (NIR). Select 589 nm for conventional analysis using chemicals or lead clarifiers, or select 880 nm for polarimetric sugar measurements in the near infrared (NIR). 880 nm is above the absorption level of most dark sugar factory solutions allowing measurement of filtered but non-lead clarified juices. The dark sample measurement bar gets darker in proportion to the darkness of the sample being measured. When the bar is completely dark the instrument has reached its measurement limit.

Dilution and Cell Length Correction

Whether your sugar analysis requires a 25mm, 50mm, 100 mm or 200 mm cell, whether you use 1/2 normal (13g/100ml), 1/4 normal (6.5g/100ml) or full weight sugar solutions, the AUTOPOL®  880 and 880T saccharimeters provide the correct multiplier to display the reading as if the sample were prepared as a 1 normal solution in a 200 mm optical path-length cell.

Automatic Calibration

Calibrating an AUTOPOL® saccharimeter is quick and simple. First, place a Rudolph Quartz Control Plate calibration standard on the optical rails and enter the Quartz Plate value in the appropriate window. Second, enter the temperature to correct the measurement to either the ICUMSA standard temperature of 20°C or the Tropical reference temperature of 27.5°C. Third, press 2 buttons and calibration is complete. (See Technical Bulletin 913 for more details on Rudolph Calibration Standards)

Automatic Temperature Correction

All AUTOPOL® saccharimeter models have a temperature probe in the measurement compartment. The probe can be inserted into all 14 style, 40T style, 41T style and 33 style measurement cells.
The AUTOPOL® 880 displays the temperature source and type of correction. Utilizing factory programmed temperature coefficients the AUTOPOL® 880 can make the following types of temperature corrections:

  • A 30° C sucrose sample can be measured by the temperature probe and corrected to 20°C or 27.5°C or a customer defined temperature.
  • A 29°C sucrose sample in a Type 14 center fill glass cell can be measured by the temperature probe inside the sample chamber and corrected to 20°C.
  • A Quartz Plate measured at 25°C by the chamber temperature probe and then automatically corrected to 20°C.

Rudolph knows Sugar Technologists around the world demand 24 hour operation, 7 days a week for up to 9 months out of the year. They expect 0.02 °Z reproducibility and accuracy. Rudolph Research understands these demands and requirements, because Rudolph has been serving the sugar industry since 1940. Throughout those sixty plus years, Rudolph has dedicated itself to manufacturing quality instrumentation and providing superior technical and service support, both directly and through its worldwide dealer network. This dedication is consistent with Rudolph’s over 5,000 instruments installed worldwide, with many of those instruments still in operation after 25 years of service.

20yearbanner

All Rudolph Instruments carry our exclusive 20 Commitment to service your instrument.

Rudolph Research Analytical offers a comprehensive range of Preventive Maintenance and Service Programs. Rudolph has demonstrated a commitment to its customers by keeping installed instruments operational, not only for 20 years, but in some cases for over 40 years. This long-term commitment to keeping our instruments running makes the cost of ownership, over the life of a Rudolph Instrument, one of the lowest in the laboratory market segment. At the date this document was published, there is no other laboratory instrument manufacturer guaranteeing service and technical support for 20 years. View the entire 20 Service Guarantee. 

Specifications

AUTOPOL 880 SACCHARIMETER SPECIFICATIONS
Optical Wavelengths Standard - 589nm, 880nm Optional - 587nm, 882 nm
Wavelength Selection Selectable: 2 standard, 4 optional
Measuring Mode °Z (°ISS), Optical Rotation or User Defined Scale
Measuring Range ±89.9° Arc Optical Rotation, ± 259°Z (ISS)
Resolution 0.001° Arc Optical Rotation, 0.01°Z (ISS)
Reproducibility 0.001° Arc Optical Rotation, 0.01°Z (ISS)
Accuracy 0.01°Z (ISS)
Measuring Scale °Z (ISS)
Prism Glan Thompson Calcite
Temperature Control Standard - by external Water Bath Optional 880T - TempTrol electronic heating and cooling 18°C - 35°C
Temperature Correction 18°C - 40°C
TempTrol™ Accuracy Temperature accuracy determined by external Water Bath
Temp. Probe Range 10° - 40°C
Temp. Probe Accuracy ±0.1°C
Measurement Time 12°/sec
Light Source Tungsten-halogen 6V, 20W, Avg. 2,000 hr. life
Optional Light Source LED 20-30 mA 100,000 hr. avg
Sample Chamber Accepts sample tubes up to 200mm
Data Storage 32 GB SSD
Communication Interface 1 x Serial, 1 x VGA, 1 x HDMI, 2 x Cat 5, 4 x USB ports
Calibration Automatic calibration via touchscreen
Purity Software Standard
Connection to Refractometer Standard
Display 800 x 600 pixel resolution with 400 nits of brightness
User Interface Touchscreen
Automatic Sensitivity Control Measures samples with transmittance as low as 0.01% (up to O.D. 4.0)
Input Power 100 - 240 Volts, 50/60Hz
Operating Dimensions 30” W x 17” H x 11” D 762 mm W x 431.8 mm H x 279.4 mm D
Shipping Dimensions 42"W x 24"H x 21"D 1066.8mm W x 267mm H x 432mm D
Operating Weight 85 lbs. (39 kg)
Shipping Weight 120 lbs. (54.5 kg)

Brochures

Saccharimeters

Scientific Solutions