Innova Porometer
Product Category: Battery Testing, Membranes, Pore Size Determination and Distribution, Surface Area
Industry: Batteries, Membrane Manufacture and Testing, Particle and surface science
Supplier: Poretech
Description
Model: Innova-100N / Innova-200N / Innova-500N
- Innova Porometer is a new type of Porometer, its test methods include gas-liquid replacement method (CFP) and liquid-liquid replacement method (LLP).
- For the type of liquid used for testing, the Innova Porometer has no restrictions, it can accurately sense the flow of various liquids.
- Pore size range: From 0.002um up to 500um.
- Flow detection range: 200LPM to 0.00005 sccm.
- Touch panel design, easy to operate.
According to ASTM F316-03
- Where a nontoxic wetting liquid can spontaneously fill the pores in the sample and a non reacting Gas (Liquid) can displace liquid from pores.
- First the largest pores will get emptied, as they require lower pressure than smaller pores.
- As the pressure is increased, more and more smaller pores are progressively emptied.
- The pressure and flow rate of Gas (Liquid) through the emptied pores provides the through pore distribution. The pressure at which through pores empty is inversely proportional to the pore size.
- Through pores are the pores connected from one side of material to the other side. Innova Porometer measures the pore throat diameter, which is the most constricted part of the entire pore path.
Application: Non-woven, Electrospinning, Glass fiber, Filter paper, Membrane-MF/UF/NF, Hollow fiber, Hemodialysis, Ceramic, Sintered metal, PTFE, Battery Separator, Fuel cell GDL
- Bubble point, Mean flow pore diameter, Smallest Pore Diameter
- Flow rate vs Pressure (Wet curve, Dry curve and Half dry curve)
- Cumulative and Differential Filter Flow%
- Pore size distribution vs Diameter
- Pore size distribution and Cumulative FF
- % Cumulative filter flow vs Diameter
- Gas permeability, Darcy’s Law
- Frequency distribution and Cum. FF
- Output: Raw Test, Excel, PDF
Specifications
Model | Innova-100N | Innova-200N | Innova-500N |
Test Method | Gas-Liquid replacement and Liquid-Liquid replacement | ||
Pressure range | 100 psi | 200 psi | 500 psi |
Maximum pore | 500 um | 500 um | 500 um |
Minimum pore | 10 nm | 5 nm | 2 nm |
Sample size | Dia. ≤ 60mm and Thickness upto 30mm | ||
Testing liquid | Galwet, Silwet, Porewet, IPA, Water, IBA… any non-corrosive liquid can be used. | ||
Sample Geometry | Flat Sheets, Rods, Tubes, Hollow Fibers, Cartridges… | ||
Pressure control | Automatic, 0-120psi / 0-250psi / 0-600 psi ; Resolution:000072 | ||
Pressure Regulator | Single | Duel, Auto switch | Duel, Auto switch |
Gas flow range | 0-200 L/min | 0-200 L/min | 0-200 L/min |
Flow Control | Automatic, 0-30cc / 0-100cc /0-500cc ; Resolution: 0.00025 | ||
Liquid flow range | 0.00005-10 sccm | 0.00005-10 sccm | 0.00005-10 sccm |
Hollow Fiber Test | Outside-in and Inside-out, the sample is placed directly in the standard sample chamber, no external test module is required. | ||
Sample chamber | Stainless steel, Two chamber design
Size:Φ99mm X 60mm (Depends on actual test requirements) |
||
Data transfer | External PC or Build-in Touch panel PC | ||
Operation | Automatic test mode and manual operate mode | ||
Test Type | Gas-Liquid method and Liquid-Liquid method
Pore size distribution Bubble point Permeability |
||
Pressurized gas | Clean and dry gas, such as air and nitrogen gas.
Air Compressor or Gas cylinders |
||
Environmental | Ambient Temperature: 20℃ to 35℃
Humidity: 35% to 55% |
||
Power | AC 100V-240V, 10A, 50/60Hz |
Accessories
Standard Accessories:
Main body, Adapter plate, Insert, Adapter/Insert remover, Non-Pore disc, Quickly connector, Drain hose, Testing liquid (Galwet/Silwet/Porewet) 100ml, Spanner wrenches, Grease, O-ring, USB/AC Power cable, Software USB flash drive, Inspection report, Operating manual(ENG).
Optional Accessories:
Hollow fiber testing adapter, EPOXY.